A Four-detector System For γ - γ Angular Correlation Studies
Document Type
Article
Publication Date
3-1-1983
Publication Source
Nuclear Instruments and Methods in Physics Research
Volume Number
206
Issue Number
3
First Page
397
Last Page
402
Publisher
Elsevier
ISSN
0168-9002
Abstract
A multiple detector system for γ - γ coincidence and angular correlation measurements is described. The system consists of four large coaxial Ge detectors set at fixed positions, enabling the measurement of γ - γ coincidences between any pair. Thus, six distinct angles are measured simultaneously. The performance of the system was studied with a 152Eu source. An application for γ - γ correlation studies in 146Ce is described. Systematic errors, if present, are shown to be less than 0.03 and 0.05 for A2 and A4, respectively. The extension of this system to more than four detectors is straightforward.
Recommended Citation
Repository citation: Wolf, A.; Chung, C.; Walters, W. B.; Peaslee, Graham; Gill, R. L.; Shmid, M.; Manzella, V.; Meier, E.; Stellts, M. L.; Liou, H. I.; Chrien, R. E.; and Brenner, D. S., "A Four-detector System For γ - γ Angular Correlation Studies" (1983). Faculty Publications. Paper 1120.
https://digitalcommons.hope.edu/faculty_publications/1120
Published in: Nuclear Instruments and Methods in Physics Research, Volume 206, Issue 3, March 1, 1983, pages 397-402. Copyright © 1983 Elsevier.